Analizzatore Gas

Analizzatore Gas

Gas analysis system for measuring and checking carburizing atmospheres including data logger with CH4 correction (NEW VERSION)

Analizzatore Gas Caratteristiche

Transportable gas analysis system for carburizing atmosphere

The device continuously measures the gas components CO, CO2, CH4 and H2 in carburizing atmospheres. The C-level, the soot limit and the gas constant Beta are calculated from the measured values.

A correction of the C-level calculation according to the residual CH4 content is included by default.

For comparative measurement according to CQI9, the C-level is calculated in parallel on the basis of the probe signal. A zirconium dioxide probe and a type S thermocouple can be connected for this purpose. Probe signal and temperature can also be simulated by manual input.

All measured and calculated values are shown on the display and can be logged in the device for later evaluation. The recording can be controlled manually or by timer.

Included in delivery is the PC software MPGA-Logger, with which the data can be read out, evaluated, printed or e.g. exported into the ECS format.

The new version of the gas analysis system has a touch display with a size of 7 inches (17.8 cm) and can now be operated even better.

With the internal timers, up to 4 potential-free external contacts can now be optionally switched, which can be used for control tasks, e.g. recording of several furnace systems by cyclic switching of the gas supply.

The measured data of CO, CO2, CH4 and H2 can now also be provided externally via 4-20 mA outputs.


Fields of application:

  • Gas analysis of carburizing processes, verification of endogas
  • C-level calculation according to the actual CH4 content
  • Comparative measurements according to CQI9
  • Compliant with SAE AMS 2759

 

Dati Tecnici
Measuring range carbon monoxide: 0 ... 30 Vol.% (optional 0 ... 50 Vol.%)
Measuring range carbon dioxide: 0 ... 0.5 Vol.% and 0 … 2 Vol.% with automatic changeover
Measuring range methane: 0 ... 10 Vol.% (optional 0 ... 25 Vol.%)
Measuring range hydrogen: 0 ... 60 Vol.% (optional 0 ... 100 Vol.%)
Calculated C-level: 0 ... 2 %C
Accuracy of measurement:   < ± 1 % relative errors (MBE)
Ambient temperature: 5 - 45 °C
Heating period:   approx. 30 min
Moisture content of measuring gas: 0 ... 10 % relative humidity (condensing humidity is not permitted)
Withdrawal of gas: integrated filter and flow meter
Gas flow rate: approx. 0.1 - 0.2 l/min (integrated pump)
Measuring speed: < 2 min
Interfaces: Mini-USB to read out the data, RS 232 for factory services
Dimensions:   19“ housing, 4U
Weight: approx. 10 kg
Connections: Zirconium oxide probe, thermocouples (Type K, N or S)


Scope of delivery:

  • Stylus for touch display
  • Mini-USB cable
  • Replacement filter insert
  • Potassium hydroxide (KOH) filter for CO2 zero point calibration
  • License for ECS Analysis
  • USB memory stick with PC software MPGA-Logger
  • Carrying case
  • 2 x Mating plug for the connection of thermocouple and C-probe
  • 2 x Diffusion-proof hose (3 meter)
  • Instruction by Stange Elektronik employees (excl. additional costs)


Options:

  • Differential pressure bubble-off bottle
  • Gas conditioning (Peltier gas cooler with pump and flow meter)

 

Downloads